Miele PG 8581 Dental Thermal Washer Installation Guide

Miele PG 8581 Dental Thermal Washer

Technical data sheet

  • PG 8581 Washer-disinfector
  • PG 8582 Med
  • PG 8583 Laboratory Glassware Washer

Legend:  Circled, bold-type abbreviations:
Connection required 

  Electrical connection Voltage (default)

Rated load

2 AC 208V 60 Hz

6.2 kW

2 AC 208V 60 Hz

6.2 kW

PG 8581

PG 8582

Fuse rating

Minimum cross-section

2 x 30 A

3 × AWG 10

2 x 30 A

3 × 5.2 mm²

  Length of electrical cord (UL 921) with plug approx. 5‘9“ approx. 1.8 m
  Electrical connection NEMA L6-30 NEMA L6-30
  Electrical connection Voltage (convertible)

Rated load

3 AC 208V 60 Hz

6.2 kW

3 AC 208V 60 Hz

6.2 kW

PG 8581

PG 8582

Fuse rating 3 x 20 A 3 x 20 A
  Minimum cross-section 4 × AWG 12 4 × 3.3 mm²
  Length of electrical cord (UL 921) with plug approx. 5‘9“ approx. 1.8 m
  Electrical connection NEMA L15-20 NEMA L15-20
  Electrical connection Voltage (default)

Rated load

2 AC 240V 60 Hz

6.2 kW

2 AC 240V 60 Hz

6.2 kW

(240V Option

PG8581)

Fuse rating

Minimum cross-section

2 x 30 A

3 × AWG 10

2 x 30 A

3 × 5.2 mm²

  Length of electrical cord (UL 921) with plug approx. 5‘9“ approx. 1.8 m
  Electrical connection NEMA L6-30 NEMA L6-30
  Electrical connection Voltage (convertible)

Rated load

3 AC 240V 60 Hz

6.2 kW

3 AC 240V 60 Hz

6.2 kW

(240V Option

PG8581)

Fuse rating 3 x 20 A 3 x 20 A
  Minimum cross-section 4 × AWG 12 4 × 3.3 mm²
  Length of electrical cord (UL 921) with plug approx. 5‘9“ approx. 1.8 m
  Electrical connection NEMA L15-20 NEMA L15-20
  Electrical connection Voltage (default)

Rated load

3 AC 208V 60 Hz

6.2 kW

3 AC 208V 60 Hz

6.2 kW

(PG 8583) Fuse rating 3 x 20 A 3 x 20 A
  Minimum cross-section 4 × AWG 12 4 × 3.3 mm²
  Length of electrical cord (UL 921) with plug approx. 5‘9“ approx. 1.8 m
  Electrical connection NEMA L15-20 NEMA L15-20
  Electrical connection Voltage (convertible)

Rated load

2 AC 208V 60 Hz

6.2 kW

2 AC 208V 60 Hz

6.2 kW

(PG 8583) Fuse rating 2 x 30 A 2 x 30 A
  Minimum cross-section 3 × AWG 10 3 × 5.2 mm²
  Length of electrical cord (UL 921) with plug approx. 5‘9“ approx. 1.8 m
      NEMA L6-30
  Electrical connection NEMA L6-30  

Installation plan

The machine is supplied with electrical cord and plug. The receptacle must be accessible after machine installation. For hard-wired machines, connection should be made via a suitable mains switch with all-pole isolation. The contact opening between all open contacts must be at least 3 mm wide and the mains switch must be lockable in the open position. If necessary, an equipotential bond with good contact connection must be provided in accordance with all appropriate national and local regulations.
Splash protection class IP 21

The machine is supplied with electrical cord and plug. The receptacle must be accessible after machine installation. For hard-wired machines, connection should be made via a suitable mains switch with all-pole isolation. The contact opening between all open contacts must be at least 3 mm wide and the mains switch must be lockable in the open position. If necessary, an equipotential bond with good contact connection must be provided in accordance with all appropriate national and local regulations.

Splash protection class IP 21

  Equipotential bonding and earthing There is a screw connection point at the back of the machine, to which equipotential bonding can be connected if required.

Connect equipotential bonding and protective earth conductor!

  Network/ printer connection (optional) Optional machine interface for transmission or process data output Communication module (not supplied)

Ethernet module or RS232 module (install electrical sockets near machine if required)

  Connections/installation must comply with IEC 60950.
  Cold Water Two cold water connections are required for machines with steam condenser (DK).Use supplied Y-adapter if only one cold water connection is available.
  Max. temperature 68 °F 20 °C
  Max. water hardness 73 gpg 70 °dH
  Recommended flow pressure 29 psi 200 kPa
  Min. flow pressure 14.5 psi 100 kPa
  Max. pressure 145 psi 1,000 kPa
  Throughput 2 gal/min 7.5 l/min
  Connection thread – on site (flat sealing) 3/4“ Male Garden Hose Thread 3/4“ Male Garden Hose Thread
  Length of connection hose (supplied) approx. 5‘ 7“ approx. 1.7 m
  Length of the steam condenser hose approx. 5‘ 7“ approx. 1.7 m
  Hot Water Connection Max. temperature Max. water hardness 150 °F

73 gpg

65 °C

70 °dH

  Recommended flow pressure 29 psi 200 kPa
  Min. flow pressure 5.8 psi 40 kPa
  Max. pressure 145 psi 1,000 kPa
  Throughput 2 gal/min 7.5 l/min
  Connection thread – on site (flat sealing) 3/4“ Male Garden Hose Thread 3/4“ Male Garden Hose Thread
  Length of connection hose (supplied) approx. 5‘ 7“ approx. 1.7 m
  Demineralised water (DI) Max. temperature Recommended flow pressure 150 °F

29 psi

65 °C

200 kPa

  Min. flow pressure 4.4 psi 30 kPa
  Max. pressure 145 psi 1,000 kPa
  Connection thread – on site (flat sealing) 3/4“ Male Garden Hose Thread 3/4“ Male Garden Hose Thread
  Throughput 2 gal/min 7.5 l/min
  Length of connection hose (supplied) approx. 5‘ 7“ approx. 1.7 m
 

PG 8583 only:

Model with feed pump for unpressurised demineralised water (ADP)

  Minimum flow pressure (DI unpressurized) Maximum pressure (DI unpressurized)

Machine connection nipple (outer dia. x l)

1.3 psi

8.7 psi

1/4“ x 3/16“

8.5 kPa 60 kPa

6 mm x 30 mm

Hose from supply container to machine to be fitted on site.
  Drain Water Drainage temperature 200 °F 93 °C
  Length of drain hose approx. 4’ 6” approx. 1.4 m
  Drain hose, max. drainage length approx. 13’ 1” approx. 4.0 m
  Drain pump head height from floor level, max. approx. 3’ 3” approx. 1 m
  Drain hose delivery, max. 4.2 gal/min 16 l/min
  On-site hose connector (external diameter x length) 7/8” × 1 3/16” 22 mm × 30 mm
  It is advisable to lay the drain hose in a curve min. 600 mm above floor level to the drainage point.
  External dispensing (optional)

(note machine model)

Connection for up to two dispensing units for liquid agents on the back panel.
  Machine feet Height-adjustable 0-2 3/8“ 0 – 60 mm
  Foot diameter 1 3/8“ 35 mm
  Entry point into housing 8 M 8 M
  Heat Dissipation to

Room

Heat transfer by emission

Load on removal

1,194 btu/h

1,365 btu/h

0.35 kWh/h

0.40 kWh/h

  Machine data Height 32 7/8“ 835 mm
  Height without lid 32 1/4“ 820 mm
  Width 23 5/8“ 598 mm
    23  
    5/8“  
  Depth   598 mm
    598 mm27  
    1/2“  
  Door height 24 1/2“ 622 mm
  Net weight 163 lbs 74 kg
  Dynamic floor load 1,200 N 1,200 N
  Width including transport pallet 26 3/8“ 670 mm
  Depth including transport pallet 29 3/8“ 740 mm
  Height including transport pallet 36 1/4“ 920 mm
  Set-up Conditions Ambient temperature Max. humidity 40-105 °F

80%

5-40 °C

80%

  Max. altitude above sea level 4,921‘ 1,500 m
 
It is essential to observe the installation instructions when installing the machine.

In order to reduce the risk of steam damage, the area around the machine should be limited to furniture and fittings that are designed for use in commercial environments.

Installation should only be performed by Miele Service or authorized installers in accordance with valid regulations, relevant standards and health and safety codes!

DIMENSIONS

  • AW Drain water
  • NW Network and printer connection
  • DK Cold water connection, steam condenser
  • PA Equipotential bonding
  • DS External dispensing system connection
  • VE Demineralized water connection
  • EL Electrical connection
  • WW Hot water connection
  • KW Cold water connection

  • Drain water
  • NW Network and printer connection
  • DK Cold water connection, steam condenser
  • PA Equipotential bonding
  • DS External dispensing system connection
  • VE Demineralized water connection
  • EL Electrical connection
  • WW Hot water connection
  • KW Cold water connection